IGA 6/23 Advanced Manual Troubleshooting · 44
Symptom Probable Cause Comments
are too high
section 1.5
Electrical noise in lead wires caused
by intense magnetic fields and/or
improper selection of interconnecting
cable and/or cable run in same
conduit as AC power lines
See section
Lens or window clouding up
Increase or decrease air purge
rate and/or investigate other
cause.
Improper grounding of cable shield
and/or IGA 6/23 Advanced housing (or
cooling jacket)
secure
fluctuations)
Target moves more than 95% out of
field of view. Conveyor type
applications will also cause
fluctuations as targets move by the
field of view
Maximum to be stored. Apply
t
clear
settings. See section 4.6
Target may be changing temperature
quickly. (Thermocouples do not
respond as fast as the
IGA 6/23 Advanced)
slowed down
Apply t
90
settings. See section
4.5
Target may be reflecting nearby
objects that have fluctuating
temperatures
Loose connection in wiring
intense magnetic fields, improper
selection of inter-connecting cable
Cable run into in same conduit as AC
power lines
Flame or reflections of flame may be
entering the field of view
Extremely dense steam, smoke, thick
dust, personnel or moving machinery
intermittently blocking 95% of
optical path of IGA 6/23 Advanced
uniform emissivity of large
moving targets
required
section 1.5
fails
The instrument‘s internal
temperature is higher than 61 °C
Note: The wavelength band of the IGA 6/23 reacts at low measuring temperatures
(below 130 °C) to incandescent lamps or very bright daylight (not valid for fluorescent
tube). For a correct measurement strong external light to the measured object should
be avoided.