IGA 6/23 Advanced Manual Settings / parameter descriptions · 24
Measuring object
(at 2.3 µm)
Measuring object
(at 2.3 µm)
Brass oxidized (tarnished)
Note: Emissivity e can be set from 5% to 100% in steps of 0.1%
4.4 Transmittance t
Transmittance is a parameter that can compensate for signal loss due to external windows etc.
For example, if the emissivity of the material is 0.6 and the transmittance of an additional
window is 0.9, then the product would be 0.54 which is well inside the allowed range.
The product of transmittance and emissivity (t x e) must not be less than 5%.
Note: Transmittance t can be set from 5% to 100% in steps of 0.1%
4.5 Response Time (t
90
)
The response time t
90
is the time interval for the analog output of the pyrometer to go from a
low temperature value up to 90% of the temperature step to a high value when measuring an
abrupt increase from said low to said high temperature.
The dynamic exposure time adjustment prolongs the exposure time at the lower range limit.
With the InfraWin software and the UPP Data Format commands it is possible to configure the
dynamic exposure time.
Dependency of the response time on th
°C
°C
°C
90
°C
°C
°C
90
Independently of this, the pyrometer performs a measurement every 25 µs and updates the
analog output. Slower response times can be used to achieve a constant temperature reading
for measuring objects that have rapidly fluctuating temperatures.
The response time is set using the InfraWin software or by using the UPP Data Format commands.
When the setting is set to min., the IGA 6/23 Advanced operates using a time constant of