Lumasense-technologies IGA 6/23 Advanced Manuel d'utilisateur Page 35

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 46
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 34
IGA 6/23 Advanced Manual Maintenance · 35
6 Maintenance
6.1 Cleaning the front window
Since the device does not contain parts that require regular maintenance, the only regular
maintenance required is periodic inspection of the front window for build-up of foreign
particiles. If allowed to build up, the particles can influence the energy received by the
instrument.
The IGA 6/23 Advanced window is not water soluable and can be cleaned with standard lens
tissue dampened with a commercially available glasses or camera lens cleaning solution. Use a
soft blower/brush (available in camera stores) to remove any grit on the window before you rub
the lens with lens tissue and solution.
Attention: NEVER CLEAN THE IGA 6/23 Advanced WINDOW WITH A DRY TISSUE
OF ANY KIND! The only time dry lens tissue may be used is to dry a window which has
already been cleaned with wet lens tissue.
6.2 Calibration
LumaSense calibrated your IGA 6/23 Advanced pyrometer at the factory and delivered your
instrument with a Works Certificate. Normally we advise against changing the factory set
calibration. If you believe that the calibration may have changed, perhaps because your
operating environment is severe, an approximate field (on-site) calibration is possible. You have
the choice of an on-site calibration or arranging a more precise calibration at the LumaSense
factory.
6.2.1 Laboratory Calibration
Contact LumaSense for information about calibration at the LumaSense factory. For most of our
customers who do not have large numbers of infrared thermometers in service, we recommend
that our laboratory be chosen to do calibration. When you have many infrared thermometers,
you may find it most convenient and economical to have your own calibration laboratory.
LumaSense has a variety of blackbody calibration sources including very economical ones.
Among these you may find the source most suitable for your own laboratory.
6.2.2 On-Site Calibration
See Chapter 9 Troubleshooting before attempting to perform your own calibration. The
instrument was calibrated at the factory to its original accuracy as stated.
Nearly all erroneous temperature readings are caused by application problems such as:
· Reflections: Unexpectedly high readings may be caused by the IGA 6/23 Advanced
“seeing” a reflection of another hotter source, especially if the target emissivity is less
than 0.8.
· Spot Sizes: Ensure the IGA 6/23 Advanced is using a proper focused distance and
takes into account the spot size in relation to measuring distance. The pyrometer can
measure objects at any distance. However, the object has to be bigger than or at least as
big as the spot size of the pyrometer in the measuring distance.
· Obscured window: Clean the window. Refer to the cleaning window section 6.1.
Vue de la page 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 45 46

Commentaires sur ces manuels

Pas de commentaire