Pearl Control Software Manual Troubleshooting 42
Degraded Image
Quality
If the camera had a good image but the quality degraded after some
time, the most common cause is a lack of camera calibration. To maintain
a high quality image, the camera needs to calibrate itself periodically.
However, this automatic calibration can be turned off, and often leads
degraded image quality.
The automatic calibration setting can be found in the lower-right corner
of the Video tab:
Alternatively, you can correct the image manually by navigating to
Shortcuts > Correct Image:
No Image 1. Check that the camera is powered on.
2. Check the video cable.
3. Change the video brightness and contrast to All Automatic.
4. Change the Automatic Mode Setting to Minimum Bias.
These two settings should result in a video single in a visible range.
Unable to see
Calibrate
Tab/Unable to use
Reset Camera
The Calibrate tab is hidden by default.
1. Click Help > Advanced.
2. Enter the password: engineer.
3. Click OK or press enter.