IS 6 Advanced / IGA 6 Advanced Manual Controls and Installation · 31
Note: The last reading is saved in the standard.i12 file and automatically appears in
this form upon opening Listing or Trend output.
Selecting file open with another file, the previous file will be overwritten and re-
placed by the standard.i12 file.
5.9 PC sampling rate (time interval between two
measurements)
This function sets a time interval. After
each interval, one measured value is
stored on the PC. Longer time intervals will result in
creating smaller stored file sizes. This function is mainly
used for long term measurements.
5.10 Spot size calculator
The InfraWin Spot Size Calculator
calculates the data for the non-
focused regions. To calculate data, enter the values
of aperture D, focused measurement distance a,
and focused measuring field diameter M as found
in the documentation relating to the optics
supplied with your specific pyrometer.
After entering the aperture and the main spot size,
the input of interim values calculates spot sizes in
different measuring distances of the fixed optics.
5.11 Search I/O Module
The I/O Module allows
accessories to connect to
the software and is used to trigger
measurement externally or to send a signal
(like a relay) under certain conditions.