IS 6 Advanced / IGA 6 Advanced Manual Introduction · 10
Environment
Protection Class: IP 65 IEC 60529 (value in mated condition)
Operating Position: Any
Ambient Temperature:
Note: During operation the
instruments will warm up and might
reach an intrinsic temperature of up
to 58 °C
0 to 70 °C at housing
Storage Temperature: -20 to 80 °C
Relative Humidity:
Non condensating conditions
Weight: 0.6 kg
Housing: Stainless steel
CE-Label: According to EU directives about electromagnetic immunity
Interface
Connection: 12-pin connector
Display (
in rear cover
):
LED, 4 digit matrix, 5 mm high temperature signal or measuring
distance
Parameter
s
:
Adjustable via interface:
emissivity, sub range, settings for maximum value storage, address,
baud rate, transmittance, response time t
90
, 0 to 20 mA or 4 to 20 mA
analog output range, °C/°F
Readable via interface: measured value, internal temperature of the
unit, measuring distance
Communication
Analog Output
:
Adjustable 0 to 20 mA or 4 to
20 mA, linear (via digital interface)
Digital Interf
ace
:
RS485 addressable (half
-
duplex)
Baud rate: 1200 Bd to 115.2 kBd (on request RS232, not addressable)
Maximum Value Storage
:
Built
-
in single or double storage. Clearing with adjusted time t
clear
(off; 10
ms; 50 ms; 250 ms; 1 s; 5 s; 25 s), via interface, automatically with the next
measuring object, hold-function
Electrical
Power Supply:
24 V DC ±
25%, ripple must be less than 50 mV
Power Consumption:
Max. 3 W (
incl. laser)
Load (analog output):
0 to
500
Ohm
Isolation:
Power supply, analog output
,
and digital interface are galvanically
isolated from each other
Note: The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.