M330 Manual Service 35
Top view of new Thermostat location.
Rear of blackbody
Front of blackbody
3. Move Transformer Primary tap from 230 VAC (H3) to 208 VAC (H2) as shown below if
operating on 220 VAC (only). Power now goes between H1 and H2 as shown below. Do
not change H3 to H2 if operating at 230 VAC.
Rear view of M330 with rear panel removed