M310 Manual Appendix B: Calibration Methods 33
Using the M310 to calibrate “fixed emissivity” pyrometers
For convenience, the table below shows what setpoint to use when calibrating pyrometers
(infrared thermometers) that have fixed emissivity at 0.95 (and operate at 8-14 µm wavelength).
This table assµmes the ambient and background are at 23 °C. If your pyrometer has adjustable
emissivity, do not use this table and set your pyrometer emissivity to 1.0.