MANUA LM390SBlackbody Calibration Source
M390S Manual Introduction 102.2 Technical DataThis Model Model Temperature Range[ ] M390-S 600 to 3000 °C (1110 to 5430 °F) Custom split designAc
M390S Manual Installation 113 Installation3.1 Installing the M390STo install the M390S:1. Allow the M390 to warm-up to room temperature for at leas
M390S Manual Installation 12To ensure consistent document formatting, this page was intentionally left blank.
M390S Manual Operation 134 Operation4.1 Starting UpTo start up the M390S:1. Turn on the power switch/breaker and allow 20 seconds for the system to
M390S Manual Operation 14Eurotherm controller4.2.1 Other Controller ModesIf, for some reason, the power must be turned off while monitoring tempera
M390S Manual Operation 159. Press the PAGE button until ACCESS appears.10. Press the SCROLL button.11. Press the DOWN button until LEVEL 1 appears
M390S Manual Operation 164.5.1 RS485 OptionRS485 two wire is available as an option. The connection scheme is shown below:RS485 for multi-drop app
M390S Manual Operation 174.6 PID Tuning ConstantsThe PID tuning constants have been factory set for optimum performance and are noted below.Heating
M390S Manual Operation 18Typical Radiance temperature calibration look-up table (LUT) usedTo gain access to the custom look-up table:Refer to the 4
M390S Manual Operation 19Note: Calibration points do not have to be used as pre-programmed. To dial in adifferent calibration point, simply change
Confidential InformationThe material contained herein consists of information that is the property of LumaSenseTechnologies and intended solely for us
M390S Manual Operation 20To ensure consistent document formatting, this page was intentionally left blank.
M390S Manual Maintenance 215 Maintenance5.1 Cavity LifeThe life expectancy of the graphite cavity depends on the time and temperature of operation.
M390S Manual Maintenance 225.4 Furnace Core DisassemblyThe graphite core can be replaced without disconnecting cooling or purge tubing. Refer toexp
M390S Manual Maintenance 23Refer to the Appendix for an exploded view of the cavity assembly.5.5 Fuse LocationsThe M390 has several fuses. Refer t
M390S Manual Maintenance 245.6 CleaningClean the outside of the cabinet with mild soapy solution and water only. Do not attempt toclean the emitte
M390S Manual Index 25IndexCCavity life 21Cleaning 24Controller Calibration 17Controller Modes 14DDisposal 7EEurotherm iTools Software 17FFurnace Co
M390 Manual Index 26To ensure consistent document formatting, this page was intentionally left blank.
M390 Manual Appendix A: Drawings 27Appendix A: Drawings
M390 Manual Appendix A: Drawings 28
M390 Manual Appendix A: Drawings 29
M390S Manual Contents iiiContents1 General Information ...
M390 Manual Appendix B: ASCII/BASIC 30Appendix B: ASCII/BASICTypical ASCII request string to read the blackbody temperature:'-------make a req
M390 Manual Appendix B: ASCII/BASIC 31A simple BASIC program to send blackbody setpoint:CLS'-----open comm port 1 for 9600 baud, 7,E,1-----OPE
M390 Manual Appendix C: Declaration of Conformance 32Appendix C: Declaration of Conformance
M390S Manual Contents ivAppendix A: Drawings ...
M390S Manual General Information 51 General Information1.1 Information about the user manualCongratulations on choosing the high quality and highly
M390S Manual General Information 6Warning: DO NOT USE NITROGEN (N2) FOR PURGE GAS. AT HIGH TEMPERATURES,CYNIDE GAS COULD BE GENERATED.1.3 Limit of
M390S Manual General Information 7Return the instrument upon receipt of the RMA nµmber, transportation prepaid. Clearlyindicate the assigned RMA nµ
M390S Manual General Information 8To ensure consistent document formatting, this page was intentionally left blank.
M390S Manual Introduction 92 Introduction2.1 DescriptionThe M390 series are high temperature blackbody calibrators that can produce a very hightemp